User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Simpson 260 5kV:Volt-Ohm-Milliammeter DATE: 14-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 132 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R F W 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP Lay the Simpson on its back for this calibration. 1.005 DISP Zero the Meter Movement using the Center Screw. 1.006 HEAD {* DC VOLTS TEST *} 1.007 DISP Connect the Red Test Lead to the (+) input terminal 1.007 DISP of the Simpson. Connect the Black Test Lead to the 1.007 DISP COMMON - Terminal of the Simpson. 1.007 DISP 1.007 DISP Connect the other end of the Test Leads to the 1.007 DISP 5700A OUTPUT TERMINALS. 1.008 DISP Set the Function Selector switch on the left hand side 1.008 DISP of the Test Instrument to + D.C. 1.009 DISP Set the Test Instrument Range to the 2.5V Range. 1.010 5700 2.5 2.50V 3% 2W #! Test Tol 0.075, Sys Tol 1.65e-005, TUR 4545.45 (>= 4.00). 2.001 ASK+ R 2.002 5700 10 1.0V 3/ 2W #! Test Tol 0.3, Sys Tol 7.2e-006, TUR 41666.67 (>= 4.00). 3.001 5700 10 5.0V 3/ 2W #! Test Tol 0.3, Sys Tol 2.9e-005, TUR 10344.83 (>= 4.00). 4.001 5700 10 10.0V 3% 2W #! Test Tol 0.3, Sys Tol 5.4e-005, TUR 5555.56 (>= 4.00). 5.001 5700 50 50V 3% 2W #! Test Tol 1.5, Sys Tol 0.0004, TUR 3750.00 (>= 4.00). 6.001 5700 250 250V 3% 2W #! Test Tol 7.5, Sys Tol 0.0026, TUR 2884.62 (>= 4.00). 7.001 5700 1000 1000V 3% 2W #! Test Tol 30, Sys Tol 0.0086, TUR 3488.37 (>= 4.00). 8.001 ASK- R 8.002 DISP Move the RED test lead to the D.C. 5000V input terminal 8.002 DISP at the lower right of the Simpson. 8.003 5700 5000 1000V 3/ 2W #! Test Tol 150, Sys Tol 0.0086, TUR 17441.86 (>= 4.00). 9.001 HEAD {* AC VOLTS TEST *} 9.002 DISP Set the Test Instrument Function switch to A.C. 9.003 DISP Move the RED test lead to the A.C. 5000V input terminal 9.003 DISP at the lower right of the Simpson. 9.004 5700 5000 1000V 5/ 400H 2W #! Test Tol 250, Sys Tol 0.084, TUR 2976.19 (>= 4.00). 10.001 DISP Move the RED test lead from the 5000V input terminal 10.001 DISP to the + terminal near the bottom left corner of 10.001 DISP the meter. 10.002 5700 1000 1000V 5% 400H 2W #! Test Tol 50, Sys Tol 0.084, TUR 595.238 (>= 4.00). 11.001 ASK+ R 11.002 5700 250 250V 5% 400H 2W #! Test Tol 12.5, Sys Tol 0.024, TUR 520.833 (>= 4.00). 12.001 5700 50 50V 5% 400H 2W #! Test Tol 2.5, Sys Tol 0.005, TUR 500.000 (>= 4.00). 13.001 5700 10 10.0V 5% 400H 2W #! Test Tol 0.5, Sys Tol 0.00082, TUR 609.756 (>= 4.00). 14.001 5700 2.5 2.50V 5% 400H 2W #! Test Tol 0.125, Sys Tol 0.0002575, TUR 485.437 (>= 4.00). 15.001 HEAD {* FREQUENCY RESPONSE TEST *} 15.002 5700 2.5 2.50V 5% 200H 2W #! Test Tol 0.125, Sys Tol 0.0002575, TUR 485.437 (>= 4.00). 16.001 5700 2.5 2.50V 5% 60H 2W #! Test Tol 0.125, Sys Tol 0.0002575, TUR 485.437 (>= 4.00). 17.001 HEAD {* DC CURRENT TEST *} 17.002 ASK- R 17.003 DISP Move the RED test lead from the + input terminal 17.003 DISP to the 50µ A terminal near the upper left corner of 17.003 DISP the meter. 17.004 DISP Set Test Instrument Function switch to D.C. 17.005 DISP Set the Test Instrument Range to the 50V(µ AMPS.) Range. 17.006 5700 50 50uA 3% 2W #! Test Tol 1.5e-006, Sys Tol 1.25e-008, TUR 120.000 (>= 4.00). 18.001 DISP Move the RED test lead from the 50µ A terminal 18.001 DISP to the + input terminal. 18.002 ASK+ R 18.003 5700 1 1.00mA 3% 2W #! Test Tol 3e-005, Sys Tol 6e-008, TUR 500.000 (>= 4.00). 19.001 5700 10 10.0mA 3% 2W #! Test Tol 0.0003, Sys Tol 6e-007, TUR 500.000 (>= 4.00). 20.001 5700 100 100mA 3% 2W #! Test Tol 0.003, Sys Tol 7e-006, TUR 428.571 (>= 4.00). 21.001 5700 500 500mA 3% 2W #! Test Tol 0.015, Sys Tol 7e-005, TUR 214.286 (>= 4.00). 22.001 DISP Set the Test Instrument Range to 10MA (amps). 22.002 DISP Move the Red wire from the meters + 22.002 DISP Input to the +10A Input and the Black lead to the 22.002 DISP -10A Input of the meter. 22.003 5700 10 10.0A 3% B1 2W #! Test Tol 0.3, Sys Tol 0.00388, TUR 77.3196 (>= 4.00). 23.001 HEAD {* RESISTANCE TEST *} 23.002 DISP Move the Red wire from the Test Instruments +10A Input 23.002 DISP to the + Input near the bottom left hand corner. 23.002 DISP Move the Black wire from the -10A Input to the 23.002 DISP - COMMON Input Terminal of the Simpson. 23.003 DISP Set the Range to {Rx1}. 23.004 DISP Short the wires from the Test Instrument together at 23.004 DISP the Test Interface Panel. 23.004 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 23.004 DISP Remove the short. 23.005 ASK+ K 23.006 5700 10.00Z +12.5% -11% 2W #! T.U.R. not calculated because tolerance or floor unspecified. 24.001 DISP Set the Range to {Rx100}. 24.002 DISP Short the wires from the Test Instrument together at 24.002 DISP the Test Interface Panel. 24.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 24.002 DISP Remove the short. 24.003 5700 1000Z +8% -7.5% 2W #! T.U.R. not calculated because tolerance or floor unspecified. 25.001 DISP Set the Range to {Rx10,000}. 25.002 DISP Short the wires from the Test Instrument together at 25.002 DISP the Test Interface Panel. 25.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 25.002 DISP Remove the short. 25.003 5700 100000Z +8% -7.5% 2W #! Test Tol 7500, Sys Tol 1.3, TUR 5769.23 (>= 4.00). 26.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 2 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.